Negotiable
The date of payment from buyers deliver within days
Beijing
Long-term effective
2020-11-20 02:33
281
Company Profile
Full Medical Co., LtdBy certification [File Integrity]
Contact:hffumai(Mr.)
Email:
Telephone:
Phone:
Area:Beijing
Address:Beijing
Website:http://www.fullmedi.com/ http://hffumai.dachaoele.com/
Products Description
* The Ventilator adopts microcomputer control with high-definition 5.7” LCD display, integrating the functions of respiratory control and displaying the monitoring parameters. It is versatile in function and excellent in performance.
* The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose two kinds of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.
* High-definition four-tube flowmeter, designed with stoppers for oxygen and nitrous oxide to control the output oxygen concentration.
* Reliable pneumatic transmission system.
* Economic and proper configuration of technologies, optional as required by the Customer.
* Modularized design and upgrading. Upgrading is available as required by the Customer.
Technical Parameters
Part of Hospital ICU Anesthesia Machine
* Main body | High-strength engineering plastic rack, light, beautiful and corrosion resistant |
* Scope of application | adult, child |
* Gas source | O2: 0.27~0.55MPa; N2O: 0.27~0.55MPa |
* Flow meter |
O2: 0.1~1.0L/min 1.1~10L/min N2O: 0.1~1.0L/min 1.1~10L/mi |
* O2, N2O linkage and N2O Stopper |
When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off. |
* Flow rate of rapid oxygen supply | 25~75L/min |
* Low oxygen pressure alarm | There will be sound alarm when the oxygen pressure < 0.2MPa |
* Vaporizer | It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. |
* Respiratory circuit |
working mode: all-close, semi-close, semi-open releasing pressure: 0.5~7 kPa |
* Respiratory bellows |
bellows for adults, bellows for children tidal volume range:0~1,500 ml |
Part of Ventilator
* Display mode | High-definition 5.7” LCD screen display | |
* Ventilation mode | IPPV,SIPPV,SIMV,PEE,MANUAL,SIGH | |
* Ventilation parameters | Tidal volume | 50~1500ml |
Rate | 2~99bmp | |
SIMV rate | 2~20bpm | |
I:E | 2:1~1:8 | |
Inspiratory trigger pressure | -1.0~2.0 kPa | |
PEEP | 0~2.0 kPa | |
Pressure range | 1.0~6.0 kPa | |
SIGH | 1.5 times the inspiratory time | |
* Parameters for ventilation monitoring |
tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure | |
* Monitoring of oxygen concentration | (21%~100%) | |
* Security Alarm System | Oxygen Concentration alarm |
upper limit setting range (50%~100%) low limit setting range (15%~50%) |
Airway pressure alarm |
upper limit setting range 1.0~6.0 kPa low limit setting range 0.4~2.0 kPa |
|
Per-minute ventilation volume alarm |
upper limit setting range 3.0~30L/min low limit setting range 1.0~10L/min |
|
Sustained high-pressure alarm | it will give alarm when stress have consistently been higher than 2.5 kPa | |
Suffocation alarm | it will give sound and light alarm if there is no tidal volume input for 15 seconds | |
Power alarm | ||
* Power | AC 220V±10%; 50Hz±1Hz; (UPS, Storage battery) | |
* (Mechanical Arms) | ||
* (End-expiratory CO2 monitoring) | ||
* Items within brackets are optional. |
http://www.fullmedi.com/